Local Operation
4-15
4.
Set Slew Rate
There are 500 discrete steps in each range. Press the
key to set slew rate of
rise. Turn the Rotary knob to change the display value to 0.03A/
μs, and the VFD
displays:
:
00.030 A/
μs
Then press
key to confirm the setting, and the slew rate settings change to fall
at the same time. Turn the Rotary knob to change the display value to 0.03A/
μs, and the
VFD displays:
:
00.030A/
μs
Then press
key to confirm the setting, and the setting page change to Power
Level at the same time.
5.
Set the second Power Level for single channel module
Press the
key to select State B then the LED “B” above the key
lights
up.
Turn the Rotary knob to change the display value to 6 watts, then press
key to
confirm.
6.0000 W
6.
Review and update the values of the setting parameters
Review the values of the setting parameters by pressing
key. Update them by
set new value in their setting pages.
After completion of the data edit,
key must be pressed. Otherwise, pressing the
key, new data will not be written into the internal memory, the previous value for
the parameter is kept.
7.
Quit from editing mode
Press
to quit from editing mode. Then, the VFD display will go back to the
voltage, current, and watt measurement display mode.
DATA
ENTER
ENTER
A/B
A/B
ENTER
DATA
ENTER
DATA
EDIT
Содержание 63600 Series
Страница 1: ......
Страница 2: ......
Страница 3: ...Programmable DC Electronic Load 63600 Series Operation Programming Manual Version 2 2 July 2017...
Страница 8: ...vi...
Страница 9: ...vii...
Страница 10: ...viii...
Страница 16: ......
Страница 39: ...General Information 1 19 Module outline Unit mm 142 85 62 510 36 6 32 UNIT mm 63610 80 60 63610 80 80...
Страница 40: ......
Страница 110: ...Programmable DC Electronic Load 63600 Series Operation Programming Manual 4 34...
Страница 132: ......
Страница 252: ......