Programmable DC Electronic Load 63600 Series Operation & Programming Manual
5-42
Query Example:
CONF:VOLT:LATC?
CONFigure:VOLTage:LATCh:RESet
Type:
Channel-Specific
Description:
Resets the Von signal.
Setting Syntax:
CONFigure:VOLTage:LATCh:RESet
Setting Parameters:None.
Setting Example: CONF:VOLT:LATC:RES
Resets the Von Signal.
CONFigure:VOLTage:OFF
Type:
Channel-Specific
Description:
Set the voltage of sink current off.
Setting Syntax:
CONFigure:VOLTage:OFF<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example:
CONF:VOLT:OFF 0.5
Set Voff = 0.5V
CONF:VOLT:OFF 500mV
Set Voff = 0.5V
CONF:VOLT:OFF MAX
Set Voff = maximum value.
CONF:VOLT:OFF MIN
Set Voff = minimum value.
Query Syntax:
CONFigure:VOLTage:OFF?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Volt]
Query Example:
CONF:VOLT:OFF?
CONF:VOLT:OFF? MAX
CONF:VOLT:OFF? MIN
CONFigure:VOLTage:ON
Type:
Channel-Specific
Description:
Set the voltage of sink current on.
Setting Syntax:
CONFigure:VOLTage:ON<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example:
CONF:VOLT:ON 0.5
Set Von = 0.5V
CONF:VOLT:ON 500mV
Set Von = 0.5V
CONF:VOLT:ON MAX
Set Von = maximum value.
CONF:VOLT:ON MIN
Set Von = minimum value.
Query Syntax:
CONFigure:VOLTage:ON?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Volt]
Query Example:
CONF:VOLT:ON?
CONF:VOLT:ON? MAX
CONF:VOLT:ON? MIN
CONFigure:VOLTage:RANGe
Type:
Channel-Specific
Description:
Set the voltage measurement range in CC mode.
Setting Syntax:
CONFigure:VOLTage:RANGEe<space><CRD | NR1>
Setting Parameters:<CRD | NR1>, LOW | L | 0, MIDDLE | M | 1, HIGH | H | 2
Setting Example:
CONF:VOLT:RANG HIGH
Set voltage range to High.
CONF:VOLT:RANG M
Set voltage range to Middle.
CONF:VOLT:RANG 0
Set voltage range to Low.
Query Syntax:
CONFigure:VOLTage:RANGe?
Return Parameters:<CRD>, LOW, MIDDLE, HIGH
[Unit = None]
Query Example:
CONF:VOLT:RANG?
Содержание 63600 Series
Страница 1: ......
Страница 2: ......
Страница 3: ...Programmable DC Electronic Load 63600 Series Operation Programming Manual Version 2 2 July 2017...
Страница 8: ...vi...
Страница 9: ...vii...
Страница 10: ...viii...
Страница 16: ......
Страница 39: ...General Information 1 19 Module outline Unit mm 142 85 62 510 36 6 32 UNIT mm 63610 80 60 63610 80 80...
Страница 40: ......
Страница 110: ...Programmable DC Electronic Load 63600 Series Operation Programming Manual 4 34...
Страница 132: ......
Страница 252: ......