Remote Operation
5-51
maximum value of static load.
CURR:STAT:FALL MIN
Set falling slew rate to the
minimum value of static load.
Query Syntax:
CURRent:STATic:FALL?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = A/
μs]
Query Example:
CURR:STAT:FALL?
CURR:STAT:FALL? MAX
CURR:STAT:FALL? MIN
CURRent:STATic:L1
Type:
Channel-Specific
Description:
Set the static load current for constant current static mode.
Setting Syntax:
CURRent:STATic:L1<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: CURR:STAT:L1 20
Set the static load parameter L1 = 20A.
CURR:STAT:L1 10A
Set the static load parameter L1 = 10A.
CURR:STAT:L1 MAX
Set the static load parameter L1 =
maximum value.
CURR:STAT:L1 MIN
Set the static load parameter L1 =
minimum value.
Query Syntax:
CURRent:STATic:L1?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Ampere]
Query Example:
CURR:STAT:L1?
CURR:STAT:L1? MAX
CURR:STAT:L1? MIN
CURRent:STATic:L2
Type:
Channel-Specific
Description:
Set the static load current for constant current static mode.
Setting Syntax:
CURRent:STATic:L2<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: CURR:STAT:L2 20
Set the static load parameter L2 = 20A.
CURR:STAT:L2 10A
Set the static load parameter L2 = 10A.
CURR:STAT:L2 MAX
Set the static load parameter L2 =
maximum value.
CURR:STAT:L2 MIN
Set the static load parameter L2 =
minimum value.
Query Syntax:
CURRent:STATic:L2?[<space><MAX | MIN>]
Return Parameters:<NR2>, [Unit = Ampere]
Query Example:
CURR:STAT:L2?
CURR:STAT:L2? MAX
CURR:STAT:L2? MIN
CURRent:STATic:RISE
Type:
Channel-Specific
Description:
Set the rising slew rate of current for constant current static mode.
Setting Syntax:
CURRent:STATic:RISE<space><NRf+>[suffix]
Setting Parameters:Refer to respective specification for valid value range.
Setting Example: CURR:STAT:RISE 2.5
Set rising slew rate to 2.5A/
μs.
CURR:STAT:RISE 1A/
μs
Set rising slew rate to 1A/
μs.
CURR:STAT:RISE MAX
Set rising slew rate to the
maximum value of static load.
CURR:STAT:RISE MIN
Set rising slew rate to the
minimum value of static load.
Содержание 63600 Series
Страница 1: ......
Страница 2: ......
Страница 3: ...Programmable DC Electronic Load 63600 Series Operation Programming Manual Version 2 2 July 2017...
Страница 8: ...vi...
Страница 9: ...vii...
Страница 10: ...viii...
Страница 16: ......
Страница 39: ...General Information 1 19 Module outline Unit mm 142 85 62 510 36 6 32 UNIT mm 63610 80 60 63610 80 80...
Страница 40: ......
Страница 110: ...Programmable DC Electronic Load 63600 Series Operation Programming Manual 4 34...
Страница 132: ......
Страница 252: ......