Programmable DC Electronic Load 63600 Series Operation & Programming Manual
4-32
P/F_VH : 5.500 V
P/F_VL : 4.500 V
P/F_IH : ------ mA
P/F_IL : ------ mA
P/F_PH : ------ mA
P/F_PL : ------ mA
h.
Setting the Sequence P/F Delay Time
The sequence Pass/Failure delay time let you set the delay time for P/F checking
when load condition changes. The failure status of the sequence will latch when a
program is executed. It means that any failure will be memorized even when the
UUT becomes stable within the specifications later. The range of P/F delay time is
from 0 to 30s. Turn the Rotary knob to change the display value to 1, then press
to set the sequence P/F delay time for 1s. This setting value must be less
than dwell time. The default setting is 0s.
P/F_DLY : 1 s
9.
Review and update the values of the setting parameters
Review the values of the setting parameters by pressing
key. Update them by
set new value in their setting pages.
After completion of the data edit,
key must be pressed. Otherwise, pressing the
key, new data will not be written into the internal memory, the previous value for
the parameter is kept.
10. Save the setting Program
There are two ways to save all sequences. One is in Auto sequences mode, press EDIT
to edit the sequence and use the rotary under the NEXT selection to select SAVE, and
then press
. The other is to use select SAVE under the NEXT selection in the
parameter setting screen of sequence and press
.
[ P01 STEP ]
P/F_PH:0.0000W
P/F_PL:0.0000W
P/F_DLY:0.0000s
NEXT:SAVE
[ AUTO SEQUENCES]
REMAIN_SEQ:100
CLEAR_SEQ:NO
TOTAL_SEQ:0
NEXT:SAVE
ENTER
DATA
ENTER
DATA
ENTER
ENTER
Содержание 63600 Series
Страница 1: ......
Страница 2: ......
Страница 3: ...Programmable DC Electronic Load 63600 Series Operation Programming Manual Version 2 2 July 2017...
Страница 8: ...vi...
Страница 9: ...vii...
Страница 10: ...viii...
Страница 16: ......
Страница 39: ...General Information 1 19 Module outline Unit mm 142 85 62 510 36 6 32 UNIT mm 63610 80 60 63610 80 80...
Страница 40: ......
Страница 110: ...Programmable DC Electronic Load 63600 Series Operation Programming Manual 4 34...
Страница 132: ......
Страница 252: ......