Table 7-11
Model
Load Voltage
Setting/Limit
Current
DC Source
Voltage/Limit
Current
DMM(V)
Front Panel Display
Reading
Max.
Min.
63630-600-15
595V/0.5A
600V/0.2A
595.8975V 594.1025V
DMM (V)±0.21V
0.8V/0.5A
5V/0.2A
1.4004V
0.1996V DMM (V)±0.150031V
63640-150-60
140V/1A
150V/0.5A
140.0725V 139.9275V
DMM (V)±0.05V
1V/1A
150V/0.5A
1.03775V 0.96225V DMM (V)±0.01525V
63630-80-60
75V/1A
80V/0.5A
75.1175V 74.8825V
DMM (V)±0.028V
1V/1A
80V/0.5A
1.0805V
0.9195V DMM (V)±0.02002V
63610-80-20
75V/1A
80V/0.5A
75.1175V 74.8825V
DMM (V)±0.028V
1V/1A
80V/0.5A
1.0805V
0.9195V DMM (V)±0.02002V
63640-80-80
75V/1A
80V/0.5A
75.1175V 74.8825V
DMM (V)±0.028V
1V/1A
80V/0.5A
1.0805V
0.9195V DMM (V)±0.02002V
*If the voltage is incapable of loading as the value set, it can add load limit current or lower
down the limit current of DC Source.
7.3.3.2 Checking Medium Voltage Range
A. Connect the Load module, DC source, DMM and current shunt as Figure 7-1 shows.
Use DMM (V) to measure the voltage passing through the module’s input terminal. Be
careful in making connections so that contact resistance voltage drop will not affect the
readings.
B. Press
till VFD shows
CV
and press
to light up the M range LED
indicator.
C. Press
to set load voltage and press
to set limit current. The DC Source
voltage output and limit current settings are based on the voltage/current values listed in
Table 7-12.
D. Next, press
to enable the load and wait for 30 seconds to record the voltage
passing through the negative input terminal.
Table 7-12
Model
Load Voltage
Setting/Limit
Current
DC Source
Voltage/Limit
Current
DMM(V)
Front Panel Display
Reading
Max.
Min.
63630-600-15
145V/1A
150V/0.5A
145.2225V 144.7775V
DMM (V)±0.05125V
2V/1A
6V/0.5A
2.151V
1.849V DMM (V)±0.0155V
63640-150-60
70V/1A
80V/0.5A
70.0375V 69.9625V
DMM (V)±0.0255V
1V/1A
80V/0.5A
1.02025V 0.97975V DMM (V)±0.00825V
63630-80-60
15V/1A
16V/0.5A
15.0235V 14.9765V
DMM (V)±0.00535V
1V/1A
16V/0.5A
1.0165V
0.9835V DMM (V)±0.00185V
63610-80-20
15V/1A
16V/0.5A
15.0235V 14.9765V
DMM (V)±0.00535V
1V/1A
16V/0.5A
1.0165V
0.9835V DMM (V)±0.00185V
63640-80-80
15V/1A
16V/0.5A
15.0235V 14.9765V
DMM (V)±0.00535V
1V/1A
16V/0.5A
1.0165V
0.9835V DMM (V)±0.00185V
*If the voltage is incapable of loading as the value set, it can add load limit current or lower
down the limit current of DC Source.
MODE
RANGE
EDIT
DATA
LOAD
Содержание 63600 Series
Страница 1: ......
Страница 2: ......
Страница 3: ...Programmable DC Electronic Load 63600 Series Operation Programming Manual Version 2 2 July 2017...
Страница 8: ...vi...
Страница 9: ...vii...
Страница 10: ...viii...
Страница 16: ......
Страница 39: ...General Information 1 19 Module outline Unit mm 142 85 62 510 36 6 32 UNIT mm 63610 80 60 63610 80 80...
Страница 40: ......
Страница 110: ...Programmable DC Electronic Load 63600 Series Operation Programming Manual 4 34...
Страница 132: ......
Страница 252: ......