Programmable Bidirectional DC Power Supply 62000D Series
Operating and Programming Manual
3-52
Figure 3-78
8.
When the APG current
calibrations are done, tap “STOP” to exit the calibration
procedure. To save the calibration data,
tap “SAVE”, and tap “CLEAR” to delete the
calibration parameters if there is no need to save them as Figure 3-78 shows.
9.
When done, return to the
“APG” of “Interface” to set the I SOURCE SET and I LOAD
SET to None.
3.2.4.4.6 APG Current Measurement Calibration
Table 3-10 lists the hardware requirements for APG current measurement calibration.
Table 3-10
Device
Suggest Model or Capacity
DVM
HP 34401A or equivalent DVM
DC Power Supply
Any DC power supply or DC signal source that
can output 10Vdc and drive more than 100mA.
Figure 3-79 shows the wire connection for APG current measurement calibration.
Содержание 62000D Series
Страница 1: ......
Страница 2: ...Get more information by downloading Chroma ATE Solutions APP ...
Страница 8: ...vi ...
Страница 9: ...vii ...
Страница 10: ...viii ...
Страница 14: ......
Страница 18: ......
Страница 35: ...Overview 1 17 Figure 1 6 Dimension of 62000D 600 1200 1800 Models ...
Страница 36: ......
Страница 52: ......
Страница 54: ...Programmable Bidirectional DC Power Supply 62000D Series Operating and Programming Manual 3 2 Figure 3 2 ...
Страница 63: ...Manual Operation 3 11 Figure 3 13 Figure 3 14 ...
Страница 109: ...Manual Operation 3 57 _EXT ON SIGNAL EXT ON DISABLE EXT ON ENABLE Output Output Figure 3 85 ...
Страница 110: ......
Страница 117: ...Program Sequence 4 7 Figure 4 8 Figure 4 9 ...