Manual Operation
3-45
6.
When above voltage calibrations are done, tap “STOP” to exit the calibration procedure.
To save the calibration data,
tap “SAVE”, and tap “CLEAR” to delete the calibration
parameters if there is no need to save them.
Improper shunt range selection may cause damage to the current shunt.
Calibration points will be different for other models (example is for the
Model 62180D-600), please following the same calibration instructions
3.2.4.4.3 APG Voltage Output Calibration
Table 3-7 lists the hardware requirements for APG voltage output calibration.
Table 3-7
Device
Suggest Model or Capacity
DVM
HP 34401A or equivalent DVM
DC Power Supply
Any DC power supply or DC signal source that
can output 10Vdc and drive more than 100mA.
Figure 3-70 shows the wire connection for APG voltage output calibration.
Figure 3-70
When conducting the APG voltage output calibration, each calibration
point has to be entered to at least 4 significant digits (i.e. 5.007V) to
ensure the power supply accuracy after calibration.
Calibration Procedure (Model 62180D-600):
WARNING
Содержание 62000D Series
Страница 1: ......
Страница 2: ...Get more information by downloading Chroma ATE Solutions APP ...
Страница 8: ...vi ...
Страница 9: ...vii ...
Страница 10: ...viii ...
Страница 14: ......
Страница 18: ......
Страница 35: ...Overview 1 17 Figure 1 6 Dimension of 62000D 600 1200 1800 Models ...
Страница 36: ......
Страница 52: ......
Страница 54: ...Programmable Bidirectional DC Power Supply 62000D Series Operating and Programming Manual 3 2 Figure 3 2 ...
Страница 63: ...Manual Operation 3 11 Figure 3 13 Figure 3 14 ...
Страница 109: ...Manual Operation 3 57 _EXT ON SIGNAL EXT ON DISABLE EXT ON ENABLE Output Output Figure 3 85 ...
Страница 110: ......
Страница 117: ...Program Sequence 4 7 Figure 4 8 Figure 4 9 ...