Installation
2-15
When the self tests of memory, data and communication are done, the screen turns to the
MAIN page automatically as Figure 2-8 shown below:
Figure 2-8
The DC power supply internal circuit may not be able to reset if it is
powered off and on immediately. It is suggested to wait for 3 seconds
after powered off and power it on again.
Before turning on the instrument, all protective grounding terminals,
extension cord and devices must connect to earth. The hazard of
potential electric shock may occur if any interrupted grounding and could
cause injury or death.
WARNING
CAUTION