Programmable Bidirectional DC Power Supply 62000D Series
Operating and Programming Manual
3-48
Figure 3-73
When conducting the APG voltage measurement calibration, each
calibration point has to enter at least four significant digits (4.005V) to
ensure the power supply accuracy after calibration.
Calibration Procedure (Model 62180D-600):
1.
In Calibration page, tap “Host V” to select APG V MEAS and list the APG voltage
calibration items as shown in Figure 3-71.
Figure 3-74
1.
When entering into the Calibration page, be sure to check the
interface connection on the rear panel is correct.