Appendix B List of Protection
B-13
detects abnormal
circuit.
DB_FAULT(*)
PWM_CH1_WARN
PWM_CH2_WARN
PWM_CH3_WARN
The internal power
components of the
DC/DC (rear)
module are short-
circuited.
1. The power
component of a
certain phase in
the DC/DC
module is
abnormal or
damaged.
2. The power
component of a
certain phase in
the DC/DC
module detects
over current
analog.
1. Remove the UUT
and check if the
operation is
correct.
2. Check and
replace the
DC/DC (rear)
module board that
has protection
occurred.
AUX_FAULT(*)
Means the auxiliary
power of the DC/DC
(rear) module is
abnormally low.
1. The circuit
detection function
is abnormal.
2. The auxiliary
power of the
DC/DC (rear)
module is
abnormally low.
Check and replace
the DC/DC (rear)
module board that
has protection
occurred.
The protection message is marked _F(FAST) and _S(SLOW) by
transient and steady state.
The protection point varies by the measurement error, thus protection
circuits may act before reaching the protection point set.
DC/DC module is divided into DC/DC(front) module and DC/DC (rear)
module.
Содержание 62000D Series
Страница 1: ......
Страница 2: ...Get more information by downloading Chroma ATE Solutions APP ...
Страница 8: ...vi ...
Страница 9: ...vii ...
Страница 10: ...viii ...
Страница 14: ......
Страница 18: ......
Страница 35: ...Overview 1 17 Figure 1 6 Dimension of 62000D 600 1200 1800 Models ...
Страница 36: ......
Страница 52: ......
Страница 54: ...Programmable Bidirectional DC Power Supply 62000D Series Operating and Programming Manual 3 2 Figure 3 2 ...
Страница 63: ...Manual Operation 3 11 Figure 3 13 Figure 3 14 ...
Страница 109: ...Manual Operation 3 57 _EXT ON SIGNAL EXT ON DISABLE EXT ON ENABLE Output Output Figure 3 85 ...
Страница 110: ......
Страница 117: ...Program Sequence 4 7 Figure 4 8 Figure 4 9 ...