Remote Operation
5-19
6.
SOURce:CURRent:LIMit:{HIGH/LOW}
Description:
Sets the output current range.
Syntax:
SOURce:CURRent:LIMIT:HIGH <NRf+>[suffix]
SOURce:CURRent:LIMIT:LOW <NRf+>[suffix]
Parameter:
Refer to individual spec for valid numeric range.
Example:
SOUR:CURR:LIMIT:HIGH 60.0 It sets the output current
range to 60A maximum.
SOUR:CURR:LIMIT:LOW 20.0 It sets the output current
range to 20A minimum.
Query Syntax:
SOUR:CURR:LIMIT:HIGH? [{MAX/MIN}]
SOUR:CURR:LIMIT:LOW? [{MAX/MIN}]
Return Parameter:
<NRf+> [Unit=Amp]
Query Example:
SOUR:CURR:LIMIT:HIGH?
It returns the current high
limit.
SOUR:CURR:LIMIT:HIGH? MAX
It returns the max. current
can be set.
Return Example:
6.01
Note
Source and Load current share this setting.
7.
SOURce:CURRent:PROTect:HIGH
Description:
Sets the current range for over current protection.
Syntax:
SOURce:CURRent:PROTect:HIGH <NRf+>[suffix]
Parameter:
Refer to individual spec for valid numeric range.
Example:
SOUR:CURR:PROT:HIGH 60.0 It sets the high limit to 60A
for current output protection.
Query Syntax:
SOUR:CURR:PROT:HIGH? [{MAX/MIN}]
Return Parameter:
<NRf+> [Unit=Amp]
Query Example:
SOUR:CURR:PROT:HIGH?
It returns the high limit of
current protection.
SOUR:CURR:PROT:HIGH? MAX It returns the max. over
current protection.
Return Example:
6.01
8.
SOURce:CURRent:SLEW
Description:
Sets the rising or falling slew rate (amp/ms) of output current.
Syntax:
SOURce:CURRent:SLEW <NRf+>[suffix]
Parameter:
Refer to individual spec for valid numeric range.
Example:
SOUR:CURR:SLEW 0.01
It sets the output current slew
rate to 0.01 Amp/ms.
SOUR:CURR:SLEW 1.00
It sets the output current slew
rate to 1.00 Amp/ms.
Query Syntax:
SOUR:CURR:LIMIT:HIGH? [{MAX/MIN}]
SOUR:CURR:LIMIT:LOW? [{MAX/MIN}]
Return Parameter:
< NRf+> [Unit=Amp/ms]
Query Example:
SOUR:CURR:SLEW?
It returns the current slew rate.
SOUR:CURR:SLEW? MAX
It returns the max. current slew
rate.
Return Example:
1.00
Note
Source and Load current share this setting.
9.
SOURce:POWer
Содержание 62000D Series
Страница 1: ......
Страница 2: ...Get more information by downloading Chroma ATE Solutions APP ...
Страница 8: ...vi ...
Страница 9: ...vii ...
Страница 10: ...viii ...
Страница 14: ......
Страница 18: ......
Страница 35: ...Overview 1 17 Figure 1 6 Dimension of 62000D 600 1200 1800 Models ...
Страница 36: ......
Страница 52: ......
Страница 54: ...Programmable Bidirectional DC Power Supply 62000D Series Operating and Programming Manual 3 2 Figure 3 2 ...
Страница 63: ...Manual Operation 3 11 Figure 3 13 Figure 3 14 ...
Страница 109: ...Manual Operation 3 57 _EXT ON SIGNAL EXT ON DISABLE EXT ON ENABLE Output Output Figure 3 85 ...
Страница 110: ......
Страница 117: ...Program Sequence 4 7 Figure 4 8 Figure 4 9 ...