INST No INE -403-0P0
−
36
−
9.2 Self-diagnosis function
Error message for abnormal operation is displayed by the self-diagnosis function.
The same error message is displayed in the detector unit and the setting display unit.
Erro
r
No.
Item
Contents
Measures
Self-
diagnosed
abnormality
----- High limit over range Measured value is
higher than display
range.
Check calibration curve
data.
Check output by checker
plate.
----- Low limit over
range
Measured value is lower
than display range.
Check calibration curve
data.
Check output by checker
plate.
Er01 Calibration data error Calibration data is
damaged.
Clear RAM.
Er02 RAM error
Detector unit ram data is
damaged.
Clear RAM.
Er03 Motor rotation
abnormal
Motor stops or rotates
abnormally.
Exchange motor.
Er04 Lamp power supply
abnormal
Lamp power supply is
damaged.
Exchange lamp power
supply unit.
Er05 A/D abnormal
A/D is over range.
Check output by checker
plate.
Er06 Computation error
Computation is
overflow.
Check output by checker
plate.
Er07 Element temperature
abnormal
Element temperature is
abnormal.
Need to return detector
unit to CHINO.
Er10 Calibration curve
data error
Calibration curve data is
not appropriate.
Reset calibration curve
data.
Er11 Knee point line
approximation error
Absorbance x is out of
range of broken line
approximation.
Reset broken line
approximation data.
Er12 Ambient temperature
too high
Ambient temperature of
detector unit is too high
Lower ambient
temperature.
Er13 Ambient temperature
too low
Ambient temperature of
detector unit is too low
Raise temperature.
Er14 Sample temperature
too high
Sample temperature is
higher than correction
range
Widen correction range.
Er15 Sample temperature
too low
Sample temperature is
lower than correction
range
Widen correction range.
Er16 Sample low
reflection
Reflection rate of
sample is low or the
lamp is damaged.
Close distance to sample.
If the lamp is damaged,
exchange it.
Er17 Regression operation
error
Regression operation is
abnormal.
Check calibration curve
sample data.
(Note)
For the items filled with
¡
in the self-diagnosed abnormality, an alarm output for
self-diagnosed abnormality is activated from the setting display unit.
9. Inspection and maintenance