INST No INE -403-0P0
−
22
−
(1) Time constant:
If a measured value fluctuates noticeably, the signal can be set to be dull by delaying the
response of this moisture meter. Set the smoothing time (equivalent to the time constant of
analog instrument) in 0.1-second increment for less than 10 seconds and in 1-second
increment for more than 10 seconds.
(2) Hold ON/OFF:
When holding of measured value is desired, turn it ON. At this time, analog output will
also be held.
(3) Preset output ON/OFF:
By turning on the preset output, this moisture meter displays the preset moisture value and
the analog output becomes the value corresponding to the preset moisture value.
(4) Decimal place in moisture display:
The decimal place can be set for easily monitoring the moisture value measured.
(5) Detector unit number:
This is the number for identifying the detector unit for connection of multiple detector units.
This setting has been completed in [6.2 Setting of detector unit number].
(6) Communications speed, Parity, Data length, Stop bit length,
BCC enable/disable:
These are for communications with a personal computer, a sequencer or other similar
devices. These data are not necessary to set for combining this moisture meter with the
setting display unit.
(7) Sample temperature correction enable/disable:
These are used for sample temperature correction enabled.
(8) Correction input scaling L, Correction input scaling H
These are used for scaling of 4 to 20mA correction input.
(9) Weight á
In 3-wavelength operation, the ratio of 2 comparative wavelengths is determined by this
value.
(10) Calibration constant K1:
This is the calibration constant at the time of zero side calibration.
(11) Calibration constant k2:
This is the calibration constant at the time of span side calibration.