6.7 Setting of calibration curve number
Set the calibration curve number according to the
measuring object.
1) By pressing CH/. key, a numeric blinks for entering a
calibration curve number.
2) Change the numeric by key, key or key.
3) When ENT key is pressed, the numeric stops blinking
and this moisture meter becomes the measurement mode
at the calibration curve number changed.
6.8
Correction of calibration curve
If the actual moisture value is different from the moisture value being preset due to the
difference of one- line and off- line, or other causes, correct it by the correction coefficients b
0
and b
1
in [6.3 Setting of calibration curve data]. Usually, the shift correction is done with
b
0
only.
CH
DATA
01 10.0
Blinking
R ef er e nc e
[Example ]
When the measured moisture before correction is 15(%H
2
O) and actual
moisture value is 13 (%H
2
O),
Set b
0
= -2(%H
2
O) .