3.3.2 Connection with detector unit
Connect the fiber optics, as shown in the following figure, through the rubber bush attached
to the detector unit.
3.3.3 Disassemble cleaning
For accurate measurements, recommend periodical cleaning of the cover glass in a liquid
contacting part of the measuring cell.
1) Remove the fiber optics (both of light
projecting side and light receiving side),
from the measuring cell.
2) Next, remove 4 screws fixing a connecting
part and remove the connecting part from
the liquid-contacting part.
3) Remove a fitting ring by using the tool A.
Then after screwing the tool B into the
cover glass, pull out the cover glass from
the liquid-contacting part.
4) Wipe off dirt on the glass surface with a
gauze, etc.
5) Confirm that an O ring is inserted in
liquid-contacting part and push in the
cover glass to the liquid-contacting part.
6) Match the positions of E and D characters
of the connecting part and the liquid
contacting part, and fix the connecting
part.
7) After the cleaning of the cover glass, make
sure to re-calibrate your moisture meter.
3. Setting
O UT
I N
Connecting part
4 - Fixing scew
Liquid contacting
part
Tool A
Tool B
Fitting ring
Curler
Cover glass
O ring
Connector for fiber
[Light projecting side ]
Connector for fiber
[Light receiving side]
Dent department
Protrusion department
D
E
Rubber bush
Bush fitting ring
Connector ring
Optical fiber
D:Connect to light projecting side
E:Connect to light receiving side