- 104 -
9-2. Recording Posotion (time) Adjustment
When the time-axis synchronization (POC) is programmed to
“ON”, the gaps between the pens may change over
time, resulting in error between their time-axis. “Recording
Position (time) Adjustment” is for correcting these changes.
(1) Pressing
the
key displays the
menu window (list of setting items).
(2) Select
“TmAxAdj”.
(3) Set pen No. to correct (adjust) with the
▲
/
▼
keys. For four pen type, each time pressing
the
▼
key changes Pen1
→
Pen2
→
Pen3
→
Pen4. Press
key to start
adjustment.
Adjustment values currently set are shown
on this window. These values are replaced
by new values, when the
key is
pressed after completing the adjustments.
(4) At first, reference pen record a straight line
on the chart paper.
Pressing
the
▼
key feeds the chart and
moves cartridge pen little to the right. A
cross point of a straight line recorded by a
reference pen and a trajectory of the
cartridge pen is adjustment ending point.
Pressing
(5) After completing the adjustment of this item,
press
key to register the recording
position.
(6) Press
key to complete pen
recording position adjustment (time).
Adjustment procedure
(1) Reference pen records a
straight line.
(3) When these two lines
match, press
key
(2) Initial position of the pen adjusted.
Display
Reference pen
Pen to be adjusted
Pen1 Plotter
pen
1
st
pen
Pen2 1
st
pen
2
nd
pen
Pen3 1
st
pen
3
rd
pen
Pen4 1
st
pen
4
th
pen
Adjusting pen selection
Ref 2
Pens stand by at 0% position.
Pen other than pen No. being adjusted
Ref 1