48
Vectra® Neo Clinical Therapy System
All waveforms in the Vectra® Neo Clinical Therapy System
are set up and edited in the same basic fashion. The
Vectra® Neo Clinical Therapy System has an Electrotherapy
Icon with the following waveforms: IFC (Interferential)
Traditional (4 Pole), TENS- Asymmetrical Biphasic,
TENS- Symmetrical Biphasic, TENS HAN, High Voltage
Pulsed Current (HVPC), VMS™, VMS™Burst, VMS™ FR,
IFC Premodulated (Traditional 2 Pole), Russian, and
Microcurrent.
Complete the following steps to begin Electrotherapy
treatment:
1. Prepare patient and therapy system for Electrotherapy.
Refer to the ELECTROTHERAPY PATIENT PREPARATION
section on page 38 for electrode selection,
preparing the patient, and securing electrodes.
2. Touch the screen to activate the system. Read and
carefully follow the instructions on the screen.
3. From the Home screen, press the Electrotherapy icon.
4. Select desired waveform from the listing on the
screen by pressing the appropriate icon. Refer to the
SYSTEM SPECIFICATIONS section on page 28 for all
waveform specifications for the Vectra® Neo Clinical
Therapy System. The screen shown below will appear.
ELECTROTHERAPY OPERATION
• To view information explaining the waveform, press
the Modality Description icon. Press the Up and
Down icons to view additional text. Press the Back
Arrow icon to return to the previous screen or the
Home icon to return to the Home screen.
• To view the most commonly used electrode
placement for the selected waveform, press the
Electrode Placement icon. A picture will appear
prompting you to select the body part you wish to
treat. Press the Down icon to scroll through text.
Press the Back icon to return to the previous screen
or the Home icon to return to the Home screen.
• To customize waveform settings shown in the
list box, press the Customize icon located in the
list box, and the screen shown below will appear.
Make the desired changes and press the Back icon
to return to the previous screen, the Home icon
to return to the home screen, or the Save icon to
save the customized settings. Refer to the CUSTOM
PROTOCOLS section on page 59 for detailed
instructions for saving customized settings.
• To set pain information for the treatment, press the
Pain Information icon. At this screen, you can do the
following:
- Map electrode placement by dragging the
electrodes to their appropriate locations on the
body and selecting either front, back, left, or
right side.
- Map pain information by touching the body area
in the list box.
- Edit pain scale information by pressing the Edit
Pain Scale icon. Press the Up and Down arrows
until pain is properly gauged. Press the Back
Arrow icon to return to the previous screen or the
Home icon to return to the Home screen.
DEVICE USER INTERFACE
Содержание Vectra Neo
Страница 1: ...Vectra Neo Clinical Therapy System User Manual Operator and Installation Instructions Rx Only ...
Страница 2: ......
Страница 5: ......
Страница 85: ......