Vectra® Neo Clinical Therapy System
GENERAL WARNINGS AND PRECAUTIONS
• Stimulus delivered by the TENS waveforms of this device,
in certain configurations, will deliver a charge of
25 microcoulombs (μC) or greater per pulse and may be
sufficient to cause electrocution. Electrical current of this
magnitude must not flow through the thorax because it
may cause a cardiac arrhythmia.
• Patients with an implanted neurostimulation device must
not be treated with or be in close proximity to any
shortwave diathermy, therapeutic ultrasound diathermy,
or laser diathermy anywhere on their body. Energy from
diathermy (shortwave, microwave, ultrasound, and laser)
can be transferred through the implanted
neurostimulation system, can cause tissue damage, and
can result in severe injury or death. Injury, damage, or
death can occur during diathermy therapy even if the
implanted neurostimulation system is turned off.
• Handle, clean, and dispose of components and accessories
that have come in contact with bodily fluids according to
National, Local, and Facility rules, regulations, and
procedures.
• This unit is considered to be a Class 3B Laser product and
thus emits visible and invisible Laser radiation (IR). Avoid
direct eye exposure to the Laser beam. The symbol to the
left is located on the back of the applicator and indicates
the active radiant surface (the area on the applicator that
emits infrared Laser energy and the direction of the beam
of light). When the unit is on, not all wavelengths are
visible to the naked eye. Therefore, when performing any
operational or functional check, always wear Chattanooga
laser protective eyewear.
• The solvents of adhesives and flammable solutions used
for cleaning and disinfecting should be allowed to
evaporate before the unit is used.
• DO NOT connect the unit to an electrical supply without
first verifying that the power supply is the correct voltage.
Incorrect voltage may cause unit damage, malfunction,
electrical shock, fire, or personal injury. Your unit was
constructed to operate only on the electrical voltage
specified on the Voltage Rating and Serial Number Plate.
Contact your DJO dealer if the unit is not properly rated.
• Laser protective eyewear should be worn during laser
treatment by the operator and patient to block infrared
light energy from the eyes during treatment.
• DO NOT point the Laser beam directly into human or
animal eyes. The lens of the eye does not detect the
invisible, coherent Laser beams, potentially resulting in
permanent retinal damage.
• Device is not designed to be used in oxygen rich
environment. Explosion hazard if the device is used in the
presence of flammable anesthetic mixture with air,
oxygen, or nitrous oxide.
DANGER