26
Vectra® Neo Clinical Therapy System
SETUP INSTRUCTIONS
• Disconnect the system from the power source before attempting
any maintenance, installation, removal, or replacement
procedures to prevent electrical shock and possible damage to
system.
• The Laser Interlock must be installed by a professional or qualified
electrician. Serious eye injury can result if the device is not
properly installed. Also, when installing the device for multiple
doors, the resistance total may not exceed 4800 ohm.
WARNING
INSTALLING THE LASER INTERLOCK
(DOOR INTERRUPT SWITCH)
The Laser Interlock is an optional safety device designed
to interrupt Laser therapy if the door to the therapy room
is opened. The Laser Interlock Kit consists of a switch
resistor and a jack. Customers must supply the necessary
cable that complies with local and international codes.
Use only qualifi ed electricians to install the Laser Interlock
Kit.
The diagrams to the right provide installation guidelines
for therapy room with single and multiple doors.
Operation of the Laser Interlock
Laser Interlock works as an interrupt switch once it is
installed and connected to the Vectra® Neo Clinical
Therapy System with Laser module.
Laser Interlock monitors the state of the door(s) of the
therapy room and only allows start of Laser treatment if all
of the doors are closed.
If any door is open, it will not allow user to start the
Laser treatment; and if treatment is already started and
someone opens the door, it interrupts the system to stop
the Laser treatment.
Diagram for Therapy Room with One Door
Diagram for Therapy Room with Multiple Doors