29
Vectra® Neo Clinical Therapy System
ULTRASOUND SPECIFICATIONS
F
requency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 MHz, ±5%; 3.3 Mhz, ±5%
Duty Cycles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10%, 20%, 50%, Continuous
Pulse Repetition Rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100 Hz ±20%
Pulse Duration . . . . . . . . . . . . . . . . . .1 mSec, ±20%; 2 mSec, ±20%; 5 mSec, ±20%
Output Power
10 cm
2
Crystal. . . . . . . . . . . . . . . . . . . . . . . 0-15 W at 1 MHz, 0-10 W at 3.3 MHz
5 cm
2
Crystal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-6W @ 1 and 3.3 MHz
2 cm
2
Crystal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-3 W @ 1 and 3.3 MHz
1 cm
2
Crystal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-1.5 W @ 3.3 MHz
Amplitude. . . . . . . . . . . . . . . . . . .0-2.5 W/cm
2
in continuous and pulsed modes
Output accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ±20%, 10% of maximum
Temporal Peak to Average Ratio: . . . . . . . . . . . . . 2:1, ±20%, at 50% Duty Cycle
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5:1, ±20%, at 20% Duty Cycle
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9:1, ±20%, at 10% Duty Cycle
Beam Nonuniformity Ratio. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5:1 maximum
Beam Type. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Collimating
IPXX Rating for Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IPX0
IPXX Rating for Applicator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IPX7
Treatment Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 to 30 min
Eff ective Radiating Areas (ERA)
ERA High
ERA Low
Description
ERA (cm
2
) cm
2
%
cm
2
%
10 cm
2
Crystal
8.5
10
+18%
7
-18%
5 cm
2
Crystal
4
5
+25%
3
-25%
2 cm
2
Crystal
1.8
2
+11%
1.4
-22%
1 cm
2
Crystal
0.9
1
+11%
0.4
-55%
Head Warming Feature
The Head Warming feature of an Vectra® Neo Clinical Therapy System utilizes
Ultrasound output, resulting in warming of the Applicator to increase patient
comfort.
With Head Warming enabled, ultrasound is emitted without pressing the Start
button while an ultrasound treatment is being set up. The Applicator LED will
not illuminate during the Head Warming period. US Channel will indicate
“Head Warming”.
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0-50% Cycling of maximum power
Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3 Mhz
Applicator Temperature. . . . . . . . . . . . . . . . . . . 29.4 °C - 43.3 °C (85 °F - 110 °F)
SYSTEM SPECIFICATIONS
The following charts represent the distribution of the
ultrasonic radiation fi eld and the orientation of the fi eld
with respect to each applicator (Y-plane represents
voltage in Vrms and X-plane represents applicator head
surface in 1mm resolution).
1 cm
2
Crystal (model 27333)
2 cm
2
Crystal (model 27334)
5 cm
2
Crystal (model 27335)
10cm
2
Crystal (model 27336)