Intelect® RPW
88
Immunity
Test
IEC 60601
Test Level
Compliance Level
Electromagnetic Environment - Guidance
Portable and mobile RF communications equipment should be used no closer to any part of the Intelect RPW, including cables, than the
recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance:
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
d = [3.5]
√
P
V
1
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
d = [3.5]
√
P
80 MHz to 800 MHz
E
1
d = [7]
√
P
800 MHz to 2.5 GHz
E
1
where
P
is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and
d
is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey
a
, should be less than the compliance level in each
frequency range
b
.
Interference may occur in the vicinity of equipment marked with the following symbol:
TABLE 2: GUIDANCE AND MANUFACTURER’S DECLARATION–ELECTROMAGNETIC IMMUNITY (CONTINUED)
The Intelect RPW unit is intended for use in the electromagnetic environment specified in the table below. The user of the Intelect RPW
unit should assure that it is used in such an environment.
NOTE 1:
At
80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast
and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the Intelect RPW is used exceeds the applicable RF compliance level above, the Intelect RPW
should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Intelect
RPW.
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.