Intelect® RPW
70
The applicator is placed over the previously diagnosed pain region.
To start and stop shock wave application, the trigger button
provided on the applicator must be pressed
Keep in mind the following:
•
Do not apply more than 300 pulses to the same spot.
•
Avoid pressing the transmitter into the treatment area with
excessive pressure. Excessive pressure is not needed for
effective and successful treatment.
•
Stop treatment after a maximum of 6000 pulses.
Positioning the Handpiece
Position the handpiece on the part of the body to be treated
according to the medical indication.
Holding the Handpiece as a Pen
The handpiece is held
behind the rear cap.
This way of holding the
applicator is recommended
for tactile treatment.
Holding the Handpiece as a Bar
The handpiece is grasped
with the entire hand
behind the rear cap.
This way of holding the
applicator is recommended
for the treatment of pain
regions covering relatively
large areas.
TREATMENT TIPS
WARNING
Operating the unit at pressures higher than 3 bar without an impact
surface can result in damage to the handset.