Intelect® RPW
82
Safety Inspections
The following safety inspections must be performed on this unit.
This must be done by persons who, based on training, knowledge
or practical experience, are capable of conducting the inspections
correctly and independently.
Visual Inspection (Daily)
When performing daily inspections of the unit, pay particular
attention the following areas of potential damage:
•
Deformation of unit housing
•
Power cable damage
•
Handpiece connection damage
•
Handpiece cable damage
Functional Test
(Daily)
When performing daily inspections of the unit, pay particular
attention the following areas of potential damage:
•
Correct function of indicators
•
Display of operating modes
NOTE:
It is the responsibility of the health care facility to
verify that the unit complies with the facility, local, and
national Earth Leakage limits.
MAINTAINING THE INTELECT RPW (CONTINUED)
DANGER
•
Under no circumstances may liquid penetrate the openings
on the unit (e.g. the connecting sockets of the handset cables).
Therefore, do not use cleaning or disinfectant sprays.
•
The unit, handpiece, and cables may not be sterilized using steam
or gas.
•
Never clean the unit with abrasives, disinfectants, or solvents that
could scratch the housing or LCD or otherwise damage the unit.
•
In order to prevent electrical shock, unplug the power plug from
the power outlet before cleaning or disinfecting the unit.