Intelect® RPW
75
6 .
Wait until the unit is
completely dry before
operating it again.
7 .
It is recommended that
the transmitters be
cleaned with an ultrasound
cleaning bath designed
for disinfecting heat-
sensitive, reusable medical
instruments. If an ultrasonic
bath is unavailable, clean
parts under running water
and wipe with an alcohol-
based surface cleaner wipe,
or spray with alcohol-based
surface cleaner and wipe
with a soft, clean cloth.
8 .
When cleaning the
transmitter, push aside
the spacer ring so that
the interior cavity can be
cleaned.
9 .
Re-assemble the transmitter and handpiece by reversing
the procedures in steps 4 and 3.
MAINTAINING THE INTELECT RPW (CONTINUED)
Hand piece Maintenance
Frequent cleaning is recommended ( After everyday of usage)
Technique
Please use cleaning brush provided with hand piece kit to clean
guide tube similar to that of cleaning a gun barrel. Also, hand clean
with soft cloth projectile.
Caution:
Do not use any lubricant when cleaning tube or projectile.