Intelect® RPW
81
15 .
Insert the shaft into the
handle and turn clockwise
until snug.
16 .
Use one hand to hold the
handle and the other hand
to tighten the shaft into
the handle using the 22
mm open face wrench.
NOTE:
Tighten the shaft
so that you are
not able to loosen
the shaft with
your hand.
17 .
Reattach the rear cap by
turning it clockwise and
holding the handpiece.
18 .
Plug in the unit and
connect the handpiece
cable to the output socket
on the unit.
19 .
Reset the pulse counter
by following the steps
beginning on
page 67
.
MAINTAINING THE INTELECT RPW (CONTINUED)
DANGER
If the unit is not safe for operation, it must be repaired by a certified
service technician and the operators must be informed of the
dangers posed by the unit.