Intelect® RPW
68
USING CLINICAL RESOURCES - UNIT SETTINGS (CONTINUED)
6 .
If you choose to use a
second handpiece during
therapy, repeat steps
1-5 for the Ch2 Default
Transmitter button.
Resetting the Pulse Counter
The Pulse Counter function allows the unit to track the number of
pulses for the transmitters for maintenance purposes. After a certain
number of pulses, the projectile must be replaced. For instructions
on replacing the projectile, see
page 76
.
The pulse counter monitors the number of pulses for a transmitter
that is connected to the same socket for each therapy session. For
example, if you keep the R-SW connected to Channel 1, the pulse
counter tracks the number of pulses for the R-SW in Channel 1. The
pulse counter will also monitor the number of pulses for a different
transmitter as long as it is always connected to the same socket.
However, the unit cannot track the number of pulses if the sockets
and transmitters are swapped (i.e., the R-SW is connected to
Channel 2). Therefore, you must connect the transmitters to the
same sockets each therapy session in order for the pulse counter to
be accurate.
NOTE:
Projectiles and guide tubes should be changed after
1,000,000 pulses.
To display the pulse count, do the following:
1 .
On the Unit Settings
screen, press the Pulse
Counter button. The Pulse
Counter screen displays.
2 .
Press the Reset button
beside the desired
transmitter to reset the
counter to zero.