Remove the applicator by pulling the cable connector only. DO
NOT remove by pulling the cable.
•
To remove the cable from the applicator, make certain the power
is off. Hold the applicator while removing the cable to prevent the
applicator from dropping to the floor.
•
Observe the patient at all times during therapy.
•
Operating the unit at pressures higher than 3 bar without an
impact surface can result in damage to the handset.
•
In the event that an Error message or Warning appears beginning
with a 2 or 3, immediately stop all use of the unit and contact the
dealer or DJO for service. Errors and Warnings in these categories
indicate an internal problem with the unit that must be tested
by DJO or a Field Service Technician certified by DJO before any
further operation or use of the system. Use of a unit that indicates
an Error or Warning in these categories may pose a risk of injury to
the patient, user, or extensive internal damage to the system.
•
Empty the water bottle at least once weekly or if it is half full. For
instructions on removing and replacing the water bottle, see
page 75
.
•
Treatment should never be given in the area of metal implants.
•
Place the unit on a flat surface during treatment and storage to
prevent tipping and rolling.
•
The surface of the transmitter can become hot. Extended skin
contact can cause minor burns. Interrupt therapy after a maximum
of 6000 pulses to allow the transmitter to cool.
•
The surface of the transmitter can become hot if the unit is used
outside of the parameters listed on
page 29
. Therefore, the operator
must monitor the temperature of the transmitter during the
treatment session.
•
Do not apply more than 300 pulses to the same spot.
•
For continued protection against fire hazard, replace fuses only
with ones of the correct type and rating.