Intelect® RPW
8
CAUTION
CAUTION
•
To determine the distance of separation for all equipment operating
near the Intelect RPW, consult the EMC tables applicable to the
other equipment.
•
This equipment is to be used by, and sold to, a trained clinician only
under the prescription and supervision of a licensed practitioner.
•
Only users trained in the use of radial pressure wave therapy should
operate the unit and its applicators.
•
The Power On/Off button is considered a "soft" switch and does
not isolate the unit from mains power when the unit is turned off.
Therefore, you must disconnect the male end of the power supply
cord from the electrical outlet (mains power supply) in order to
isolate it.
•
The Intelect RPW is not intended for the treatment of kidney stones.
•
Depress the tabs of the caster locks to prevent the front wheels from
moving during treatment and storage.
•
Disconnect the power supply cord before removing covers on
this equipment. Refer the servicing of this unit to qualified service
personnel.
•
This equipment has an output that is capable of producing a
physiological effect.
•
Grounding reliability can only be achieved when the equipment is
connected to an equivalent receptacle marked "Hospital Only" or
"Hospital Grade."
•
Medical electrical equipment needs special precautions regarding
EMC and needs to be installed and put into service according to
the EMC information provided in this manual.
•
This unit can radiate radio frequency energy and, if not installed
and used in accordance with the instructions, may cause harmful
interference to other devices in the vicinity. However, there is no
guarantee that interference will not occur in a particular installation.
Harmful interference to other devices can be determined by turning
this equipment on and off. Try to correct the interference using one
or more of the following: reorient or relocate the receiving device,
increase the separation between the equipment, connect the
equipment to an outlet on a different circuit from that to which the
other device(s) are connected, and consult the factory field service
technician for help.