18
Intelect Transport® Combo Therapy System
5.5 VISUAL INSPECTION
General
Visually inspect the Intelect Transport Therapy
System. A visual inspection can, to an experienced
technician, indicate possible abuse of the unit and
internal
problems.
5.6 UNIT STARTUP AND FAN TESTING
A. Test
1. Place unit face up on work surface.
2. Connect power cord to unit and plug into
proper
power
receptacle.
3. Turn system on. Press the Enter button. IFC
should be highlighted. Press the Enter
button.
4. Place hand at the back of system, at Contrast
Control, to verify fan is blowing out.
See
Figure
5.1.
B. Test Results
1. Unit will not start, unit failed test.
a) Possible bad Main Power Switch.
b) Possible bad Power Supply.
c) Possible bad power outlet or Mains Power
Cord.
2. Screen does not display, unit failed test.
a) Contrast Control needs adjusting.
b) Possible bad display.
c) Possible bad Control Board.
d) Possible bad Power Supply.
e) Visually check power LED. LED should
illuminate Blue. Turn system off with Power
button. Power LED should flash Blue.
If Power LED illuminates Blue with system
On and flashes Blue with system Off, the
Power Supply is good. Replace Control
Board.
3. Fan not blowing outward, Unit failed test
a)
Fan
blowing
inward.
Fan wired wrong. Rewire or replace Fan.
b)
Fan
not
blowing.
1) Possible bad Fan.
2) Possible bad Power Supply.
3) Possible bad Control Board.
FIGURE 5.1
5 TROUBLESHOOTING
Содержание 2738
Страница 1: ...ISO 13485 CERTIFIED Moving Rehabilitation Forward ...
Страница 2: ......
Страница 52: ......