5 TROUBLESHOOTING
5.3
ELECTRICAL SAFETY
The Intelect Transport Combo System has been
tested to UL 60601-1, Standard for Safety for
Medical
Equipment.
A.
Power
Requirements
Model: 2738 . . . . . . . . . . . . . . . . . . . . . . Input : 100-240V
100 VA, 50/60 Hz
Listed by Intertek Testing Services NA
Inc.
Conforms to UL Standard 60601-1
Certified to CAN/CSA Standard C22.2
No. 601.1-M90 w/A2
9700675
NOTE:
NOTE:
This device complies with current leakage,
This device complies with current leakage,
ground continuity, and dielectric withstand
ground continuity, and dielectric withstand
(Hi-Pot) limits as prescribed by IEC/EN/UL
(Hi-Pot) limits as prescribed by IEC/EN/UL
60601-1 and CSA/CAN 601.1 Medical Electrical,
60601-1 and CSA/CAN 601.1 Medical Electrical,
Part 1: General Requirements for Safety.
Part 1: General Requirements for Safety.
Facility, local and national limits and test
Facility, local and national limits and test
methods may vary.
methods may vary.
UNIT FAILING DIELECTRIC WITHSTAND OR LEAKAGE
TESTS COULD INDICATE SERIOUS INTERNAL PROBLEMS.
DO NOT PLACE UNIT BACK INTO SERVICE! SEND UNIT
TO FACTORY FOR REPAIR! DO NOT ATTEMPT TO REPAIR!
5.4 LEAKAGE TESTS
Conduct all necessary leakage tests as required
per NFPA 99 (National Fire Protection Association)
“Health Care Facility” standards.