28
Intelect Transport® Combo Therapy System
5.16 COMBO OPERATION TEST
This test is performed using the 5 cm
Applicator.
Highlight Channel 1 and set up system per
5.6
parts A and B
prior to performing tests.
Connect the Intelect Transport 5 cm Applicator
to the System.
See Figure 5.18
.
Applicator LED will flash green five times.
A. Combo Operation Test Procedures
1. Highlight Combo. Press the Enter button.
2. Highlight Display. Press the Enter button
until Watts is displayed beside Display.
3. Highlight Waveform. Press the Enter button.
4. Press the Up or Down Arrow button until
IFC is highlighted. Press the Enter button.
5. Highlight Edit Stim. Press the Enter button.
Increase Intensity until Channel 1 reads
50
mA.
7. Press START button.
8. Touch the Ultrasound Applicator to the
Combo Contact on the ESTI-2 Load Test
Box. The Combo Indicator on the ESTI-2
should
illuminate.
See Figure 5.19.
B. Combo Operation Test Results
1. Combo Indicator light illuminates.
Unit passed test.
2. Combo Indicator light does not
illuminate.
Unit failed test.
Replace Channel 1 Stim Board.
FIGURE 5.19
FIGURE 5.18
COMBO INDICATOR
ILLUMINATED
5 TROUBLESHOOTING