27
Intelect Transport® Combo Therapy System
B. Ultrasound Duty Cycle Test Results
1. Duty Cycles fall within the specified ranges
as
listed
in
Figure 5.17.
Unit passed test.
2. Readings fall outside specified ranges of
Figure
5.17.
a) Possible bad degassed water in Power
Meter.
b) Possible use of Power Meter other than
Ohmic Instruments DT 100 or UPM DT 10
Ultrasound
Power
Meter.
c) Possible bad or out of calibration
Applicator. Retest with known good
Intelect
Transport
Applicator.
d) Possible bad internal connection at
Ultrasound
Board.
e) Replace possible bad Ultrasound Board.
f ) Replace possible bad Control Board.
5 TROUBLESHOOTING
5.15 ULTRASOUND DUTY CYCLE TEST CONTINUED