Intelect Transport® Combo Therapy System
6.6 ULTRASOUND BOARD
A. Part Number ............................................
27269
B. Tools and Equipment Required
#1 Phillips Screwdriver
1/4" Nut Driver or Wrench
C. Ultrasound Board
1. Separate Top from Bottom.
Refer to 6.1,
part
C.
2. Remove Channel 1 Stim Board.
Refer to 6.4,
part
C.
3. Remove Channel 2 Stim Board.
Refer to 6.5,
part
C.
4. Using the 1/4" Nut Driver, remove the four
Stand Offs securing Ultrasound Board in
place.
See Figure 6.15.
5. Carefully remove the Ultrasound Board.
See
Figure
6.16.
NOTE:
Headers may stay on the board being
removed or on the connector. If the header
stays with the connector, remove and install
on the replacement board.
6. Remove the 40 Pin Header from the
Ultrasound Board or the connector.
See
Figure
6.16.
7. Install 40 Pin Header to back of new
Ultrasound Board. Make certain it is
completely
seated
against
board.
See
Figure
6.17
.
8. Position the Ultrasound Board over the
Control Board aligning the 40 Pin Header
with the 40 Pin Connector on Control Board.
See Figure 6.16.
9. Press the Ultrasound Board into position
until the board rests against the Control
Board Stand Offs. Verify that it is well seated
by pushing on the sides of the connector.
10. Secure the Ultrasound Board using the
Stand Offs removed in
part B, step 4
above.
11. Reassemble by reversing
steps 1
through
3
.
•
•
6 REMOVAL & REPLACEMENT
FIGURE 6.15
FIGURE 6.16
Unplug the unit from the power source before
attempting removal or replacement procedures to
prevent electrical shock.
FIGURE 6.17
REMOVE STANDOFFS
CONTROL BOARD CONNECTOR
INSTALL HEADER TO BACK OF
ULTRASOUND BOARD
CONTROL BOARD HEADER
Care must be taken during removal and replacement
with the Header Pins. Bending the Header Pins may
cause the unit to fail.