Intelect Transport® Combo Therapy System
8.1 GENERAL PROCEDURES
A. Tools and Equipment Required
All Ultrasound Applicators for the unit being
serviced.
Ohmic Instruments UPM DT 10 or UPM DT
100 Ultrasound Power Meter, set to "Watts".
Degassed Water.
Refer to page 16
for
Degassed Water Recipes.
B. Ultrasound Applicator Calibration
Procedures
1. Perform the following on all Ultrasound
Applicators for the unit being serviced at
least
annually.
2. With the system on, press the Clinical
Resources
button
once.
See Figure 8.1.
3. Simultaneously press and hold the Treatment
Time and Intensity Decrease buttons for
approximately
2
seconds.
See Figure 8.2
. The
calibration procedures screen should display.
4. Press the Down Arrow button until US
Applicator Calibration is highlighted.
5. Press the Enter button.
6. Place the Ultrasound Applicator being
calibrated into the Ohmic Instruments
UPM DT 100 or DT 10 Ultrasound Power
Meter and set meter to "Watts”.
See
Figure
8.3
.
7. Follow the instructions on the LCD Display.
8. When calibration is complete, follow the
test
procedures
in
5.13 and 5.14
to verify
that the Applicator is calibrated..
•
•
•
FIGURE 8.3
FIGURE 8.1
FIGURE 8.2
SET METER
TO WATTS
8 ULTRASOUND APPLICATOR CALIBRATION
Use only Degassed Water in Power Meter for
calibrating Ultrasound Applicators.
Use of other types of water will cause false readings
and bad test results.
See page 16
for Degassed Water Recipes.
Use of other brands or types of tools, equipment,
fixtures, materials, and supplies other than those
specifically listed on
page 16
will give bad test and
calibration results.
If proper equipment is not available or cannot be
obtained, send the Ultrasound Applicators to the
factory for calibration.