Intelect Transport® Combo Therapy System
5.12 ULTRASOUND TESTS
Equipment Required for 5.13 and 5.14
Degassed Water.
Refer to page 16
for
Degassed Water Recipes.
Ohmic Instruments DT 100 UPM or DT 10
Ultrasound Power Meter.
Dissolved Oxygen Test Kit. Used to test
oxygen level of degassed water.
Intelect Transport Combo Applicator.
5.13 ULTRASOUND APPLICATOR IDENTIFICATION
TEST
NOTE
:
Use any Intelect Transport Combo Applicator for
this
test.
A. Ultrasound Applicator Identification
Test
Procedures
1. Without Ultrasound Applicator installed, turn
unit
on.
2. View the Ultrasound channel in the lower right
corner of screen. It should read “Unplugged”.
See Figure 5.11.
3. Connect Ultrasound Applicator into
Applicator
receptacle.
See Figure 5.12.
Watch
Applicator LED while connecting to system.
The LED should flash Green five times.
4. Look at the Ultrasound channel. It should
read
Available.
See Figure 5.12.
5. Highlight Ultrasound. Press the Enter button.
6. Highlight Warming. Press the Enter button
until On is displayed beside Warming.
7. Press the Back button. Turn System Off and
back On with Main Power Switch. After
System boots, view the Ultrasound channel,
Warming should be visible.
See Figure 5.13.
•
•
•
•
FIGURE 5.11
FIGURE 5.12
FIGURE 5.13