36
Intelect Transport® Combo Therapy System
6.7 CONTROL BOARD ASSEMBLY
A. Part Number ..............................................
27249
B. Tools and Equipment Required
#1 Phillips Screwdriver
1/4" Nut Driver or Wrench
C. Control Board Replacement
1. Separate Top from Bottom.
Refer to 6.1,
part
C.
2. Remove Channel 1 Stim Board.
Refer to 6.4,
part
C.
3. Remove Channel 2 Stim Board.
Refer to 6.5,
part
C.
4. Remove Ultrasound Board,
Refer to 6.6,
part
C.
5. Remove the Contrast Control Knob.
See
Figure
6.18.
6. Remove the seven screws securing the
Control Board Assembly in position.
See
Figure
6.19.
7. Remove the four LCD Bracket Mounting
Screws.
See Figure 6.21.
8. While lifting on the lower end of the Control
Board Assembly with one hand, release the
four Tabs securing the Control Board with the
other, to remove the Control Board from the
system Top. Two of the Tabs are located
above the LCD and two below.
See
Figures 6.19 and 6.20.
•
•
6 REMOVAL & REPLACEMENT
FIGURE 6.18
FIGURE 6.19
Unplug the unit from the power source before
attempting removal or replacement procedures to
prevent electrical shock.
FIGURE 6.20
REMOVE
CONTRAST KNOB
REMOVE SCREWS
REMOVE SCREWS
LIFT
HERE
RELEASE TABS
CONTROL BOARD
TABS ABOVE LCD