Intelect Transport® Combo Therapy System
1.1 PRECAUTIONARY SYMBOL DEFINITIONS
The precautionary instructions found in this manual
are indicated by specific symbols. Understand these
symbols and their definitions before operating or
servicing this equipment. The definitions of these
symbols are as follows:
A. CAUTION
Text with a “CAUTION” indicator will explain
possible safety infractions that have the
potential to cause minor to moderate injury or
damage to equipment.
B. WARNING
Text with a “WARNING” indicator will explain
possible safety infractions that will potentially
cause serious injury and equipment damage.
C. DANGER
Text with a “DANGER” indicator will explain
possible safety infractions that are imminently
hazardous situations that would result in death
or serious injury.
D. DANGEROUS VOLTAGE
Text with a “Dangerous Voltage” indicator
serves to inform the user of possible hazards
resulting in the electrical charge delivered to the
patient in certain treatment configurations of
TENS
waveforms.
E. CORROSIVE HAZARD NIMH BATTERY
Text with a “Corrosive Hazard” indicator will
explain possible safety infractions if the
chemical components of this product are
exposed to air, skin, or other materials.
F. NOTE:
Throughout this manual “NOTE” may be found.
These Notes are helpful information to aid in
the particular area or function being described.
1.2 SAFETY PRECAUTIONS
Read, understand, and follow all safety precautions
found in this manual. Below are general safety
precautions that must be read and understood before
attempting any service techniques on these systems.
Read, understand, and practice the precautionary and
operating instructions. Know the limitations and hazards
associated with using any electrical stimulation or ultrasound
device. Observe the precautionary and operational decals
placed on the unit.
DO NOT operate the Intelect Transport Combo unit when
connected to any unit other than Chattanooga Group devices.
DO NOT operate this unit in an environment where
other devices are being used that intentionally radiate
electromagnetic energy in an unshielded manner. Portable
and mobile RF communications equipment can affect Medical
Electrical Equipment.
DO NOT use sharp objects such as a pencil point or ballpoint
pen to operate the buttons on the control panel.
This unit should be operated, transported, and stored in
temperatures between 15 °C and 40 °C (59 °F and 104 °F),
with relative humidity ranging from 30%-60%, and where the
atmospheric pressure is between 950 h Pa and 1050 h Pa.
The Intelect battery pack is designed for use only with
Chattanooga Group Intelect Transport, Stim, Combo, and
Ultrasound systems.
The unit should be routinely checked before each use to
determine that all controls function normally; especially that
the intensity control properly adjusts the intensity of the
electrotherapy and ultrasonic power output in a stable manner.
Also, determine that the treatment time control actually
terminates electrotherapy and ultrasonic power output when
the timer reaches zero.
Inspect cables and connectors before each use.
The Intelect Transport Combo is not designed to prevent the
ingress of water or liquids. Ingress of water or liquids could
cause malfunction of internal components of the system and
therefore create a risk of injury to the patient.
DO NOT permit any foreign materials or liquids to enter the
unit. Take care to prevent any foreign materials including, but
not limited to, inflammables, water, and metallic objects from
entering the unit. These may cause unit damage, malfunction,
electrical shock, fire, or personal injury.
Handle the applicator with care. Inappropriate handling of the
applicator may adversely affect its characteristics.
Before each use, inspect the applicator for cracks, which may
allow the ingress of conductive fluid.
This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance
with the instructions, may cause harmful interference to other
devices in the vicinity. However, there is no guarantee that
interference will not occur in a particular installation. Harmful
interference to other devices can be determined by turning
this equipment on and off. Try to correct the interference using
one or more of the following: reorient or relocate the receiving
device, increase the separation between the equipment,
connect the equipment to an outlet on a different circuit from
that to which the other device(s) are connected and consult
the factory field service technician for help.
•
•
•
•
•
•
•
•
•
•
•
•
•