Table 6-2: Ultra-low Variable Energy (150 VE) Powerheart® G3 Waveform
Technical Data
Phase 1
Phase 2
Patient’s
Impedance
(Ohms)
Current*
(A)
Voltage*
(Volts)
Duration*
(MS)
Current*
(A)
Voltage*
(Volts)
Duration*
(MS)
Nominal
Energy**
(J)
25
56
1393
3.3
30
743
3.2
170
50
28
1420
4.5
18
909
3.2
150
75
19
1430
5.8
13
973
3.2
136
100
14
1434
7
10
1007
3.2
127
125
11
1437
8.3
8
1027
3.2
120
150
10
1439
9.5
7
1040
3.2
115
175
8
1441
10.8
6
1049
3.2
111
Table 6-3: Low Variable Energy (200 VE) Powerheart® G3 Waveform
Phase 1
Phase 2
Patient’s
Impedance
(Ohms)
Current*
(A)
Voltage*
(Volts)
Duration*
(MS)
Current*
(A)
Voltage*
(Volts)
Duration*
(MS)
Nominal
Energy**
(J)
25
64
1609
3.3
34
858
3.2
226
50
33
1640
4.5
21
1050
3.2
200
75
22
1651
5.8
15
1124
3.2
182
100
17
1656
7
12
1163
3.2
169
125
13
1660
8.3
9
1186
3.2
160
150
11
1662
9.5
8
1201
3.2
153
175
10
1663
10.8
7
1212
3.2
148