6-12
70-02053-31 A
Powerheart® AED G3 Elite 9790A and 9790E
Technical Data
Table 6-8: Second and subsequent shocks – Low Energy (200 VE)
Powerheart® G3 Pediatric Waveform
Phase 1
Phase 2
Patient’s
Impedance
(Ohms)
Voltage*
(Volts)
Duration*
(MS)
Voltage*
(Volts)
Duration*
(MS)
Nominal
Energy**
(J)
25
430
5.8
295
3.2
41
50
630
6.5
425
3.2
47
75
745
7.0
510
3.2
46
100
790
7.4
560
3.2
43
125
855
7.8
610
3.2
39
Table 6-9: Second and subsequent shocks – High Energy (300 VE)
Powerheart® G3 Pediatric Waveform
Phase 1
Phase 2
Patient’s
Impedance
(Ohms)
Voltage*
(Volts)
Duration*
(MS)
Voltage*
(Volts)
Duration*
(MS)
Nominal
Energy**
(J)
25
500
5.8
380
3.2
56
50
700
6.5
520
3.2
63
75
820
7.0
620
3.2
62
100
920
7.4
680
3.2
58
125
960
7.8
720
3.2
53
* All values are typical.
**Actual energy del/- 15%.