3-10
70-02053-31 A
Powerheart® AED G3 Elite 9790A and 9790E
Getting Started
Voice Prompt
Text display
Situation
Note:
The AED is shipped from the factory with ENHANCED MODE defaulted ON. The
Medical Director may modify the CPR options in MDLink®. ENHANCED CPR prompts
are listed in this table. Except where noted, prompts apply both to compressions-only
CPR and traditional CPR (compressions and breaths).
“It is now safe to touch
NOW SAFE
Advises the rescuer that it is safe
the patient.”
TO TOUCH
to touch the patient:
THE PATIENT
- After the AED delivers a shock
- After the AED detects a
non-shockable cardiac rhythm.
“When instructed, give patient 30 COMPRESSIONS This prompt plays at the start of a
30 rapid compressions then
2 BREATHS
CPR interval where the AED
give 2 breaths”
detects a non-shockable heart
rhythm.
Note:
Prompt for traditional CPR
only.
“Place heel of one hand on
PLACE ONE HAND Prompts rescuer to correctly
center of chest between
ON CENTER
place one hand for giving
nipples.”
OF CHEST
compressions.
“Place heel of other hand
PLACE OTHER
Prompts rescuer to correctly
directly on top of first hand.
HAND ON TOP
place other hand and body for
Lean over patient with
OF FIRST HAND
giving compressions
elbows straight.”
“Press the patient’s chest down PRESS CHEST
Prompts the rescuer to press
rapidly one third depth of
DOWN
down one third depth of patient’s
chest, then release”
FIRMLY
chest.
“Start CPR”
START CPR
Prompts to start CPR.
“Press” (30 times at 100/minute) {CPR COUNTER}
CPR counter shows the amount
of time remaining for CPR session.
(or)
Metronome (30 times at 100/minute)
(or)
No Prompt (silence)
Note:
Option is selected in MDLink
software.
Table 3-6: CPR prompts