Powerheart® AED G3 Elite 9790A and 9790E
70-02053-31 A
6-11
Table 6-5: Initial shock – Ultra Low Energy (150 VE) Powerheart® G3 Pediatric Waveform
Phase 1
Phase 2
Patient’s
Impedance
(Ohms)
Voltage*
(Volts)
Duration*
(MS)
Voltage*
(Volts)
Duration*
(MS)
Nominal
Energy**
(J)
25
370
6.1
258
3.2
31
50
550
7.3
366
3.2
36
75
640
8.6
417
3.2
37
100
705
9.8
442
3.2
36
125
770
11.1
453
3.2
35
Table 6-6: Initial shock – Low Energy (200 VE) Powerheart® G3 Pediatric Waveform
Phase 1
Phase 2
Patient’s
Impedance
(Ohms)
Voltage*
(Volts)
Duration*
(MS)
Voltage*
(Volts)
Duration*
(MS)
Nominal
Energy**
(J)
25
430
6.1
298
3.2
42
50
630
7.3
422
3.2
50
75
745
8.6
482
3.2
51
100
790
9.8
511
3.2
49
125
855
11.1
524
3.2
47
Table 6-7: Second and Subsequent Shocks – Ultra Low Energy (150 VE)
Powerheart® G3 Pediatric Waveform
Phase 1
Phase 2
Patient’s
Impedance
(Ohms)
Voltage*
(Volts)
Duration*
(MS)
Voltage*
(Volts)
Duration*
(MS)
Nominal
Energy**
(J)
25
370
5.8
270
3.2
31
50
550
6.5
390
3.2
35
75
640
7.0
470
3.2
34
100
705
7.4
510
3.2
32
125
770
7.8
545
3.2
29
Technical Data