Powerheart® AED G3 Elite 9790A and 9790E
70-02053-31 A
3-9
Voice Prompt
Text display
Situation
“Do not touch patient!
DO NOT TOUCH Repeats until analysis of the
Analyzing heart rhythm.
PATIENT
patient’s cardiac rhythm is
Please wait.”
ANALYZING
completed. This prompt will be
RHYTHM
interrupted when ready to shock.
“Preparing shock. Move
NO CONTACT
Repeats while the AED is preparing
away from the patient!”
WITH THE PATIENT to deliver a defibrillation shock
(charging).
Table 3-3: Analysis
Voice Prompt
Text display
Situation
“Press red flashing button to
PRESS BUTTON
Prompts after the AED is fully
deliver shock.”
TO DELIVER
charged and ready to deliver the
SHOCK
defibrillation shock. The RED
SHOCK indicator flashes and the
phrase repeats for 30 seconds or
until the SHOCK button is pushed.
“Shock delivered”
SHOCK
Prompts when the shock is
DELIVERED delivered.
Table 3-4: Delivering shock - Semi-automatic
Getting Started
Voice Prompt
Text display
Situation
“Shock will be delivered in”
SHOCK IN:
After the AED is fully charged and
ready to deliver the defibrillation
shock. The SHOCK will
automatically be administered
approximately three seconds
after the end of the voice prompt.
“Three”
THREE
Prompts approximately three
seconds prior to delivering shock.
“Two”
TWO
Prompts approximately two
seconds prior to delivering shock.
“One”
ONE
Prompts approximately one
second prior to delivering shock.
“Shock delivered”
SHOCK
Prompts when the shock is
DELIVERED delivered.
Table 3-5: Delivering shock - Fully automatic