Powerheart® AED G3 Elite 9790A and 9790E
70-02053-31 A
3-7
RescueCoach™ voice prompts and text display
The RescueCoach voice prompts activate when the AED lid is
opened and help guide the operator through the rescue. The AED
text display provides a visual display of most of the audible voice
prompts.
The following tables list the voice and text prompts and a
description of when the prompts are issued.
Table 3-1: Initial instructions
Voice Prompt
Text display
Situation
“Stay calm. Follow these voice
CALL 000!
Plays after lid opening self test,
instructions. Make sure 000 is
default ON.
called now!”
“Stay calm. Follow these voice
CALL EMERGENCY Medical Director may use MDLink®
instructions. Make sure
SERVICES NOW!
to select this prompt instead of
Emergency Services are called
“CALL 000!”.
now!”
MDLink also allows emergency
services and 000 prompts to be
disabled.
Table 3-2: Preparation
Voice Prompt
Text display
Situation
“Begin by exposing patient’s
BARE PATIENT’S
Prompts the rescuer to remove
bare chest and torso.
TORSO
patient clothing.
Remove or cut clothing
REMOVE
if needed.”
CLOTHING
Getting Started