Powerheart® AED G3 Elite 9790A and 9790E
70-02053-31 A
2-11
Table 2-2: Biphasic Energy Protocols (continued)
Shock
Energy Level
Energy Protocols
Sequence
1
(VE)
Energy Range
2
(J)
Protocol #4
1
150
95-196
2 150 95-196
3 200 126-260
Protocol #5
1
200
126-260
2 200 126-260
3 200 126-260
1
The Ultra-Low Energy (150 VE), Low Energy(200 VE) and High Energy(300 VE)
shocks are variable energy. The actual energy is determined by the patient’s
impedance.
2
Allowable energy range.