Powerheart® AED G3 Elite 9790A and 9790E
70-02053-31 A
5-3
Indicator troubleshooting table
The following is a troubleshooting table for the AED indicators.
Table 5-1: Indicator Troubleshooting Table
View
Symptom Solution
Red Service indicator
Maintenance by authorized
(LED) is lit.
service personnel is required.
Contact Cardiac Science
Technical Support or, outside
the U. S., your local Cardiac
Science representative.
Red Pads indicator
Connect the pads or replace
(LED) is lit.
with a new pair.
The last battery
The battery is low. Replace with
indicator (LED) is red.
a new battery.
Rescue Ready® Status
Replace the battery. If the status
indicator is red, and
indicator remains red, contact
no other indicators
Cardiac Science Technical
on the diagnostic
Support or, outside the U. S.,
panel are lit.
your local Cardiac Science
representative.
0%
100%
Troubleshooting and Maintenance