Powerheart® AED G3 Elite 9790A and 9790E
70-02053-31 A
6-3
Table 6-1: Parameters (continued)
Parameter Detail
Batteries and Capacitor
A new battery, after the AED has delivered 15 300VE
Charge Times
shocks, typically takes 10 seconds to charge the AED to
maximum energy.
A battery with reduced capacity will take longer to
charge the AED.
AED Self test Sequence
Daily: Battery, pads, internal electronics, Shock button,
and software.
Weekly: Battery, pads, internal electronics, Shock button,
software, and partial energy charge cycle.
Monthly (every 28 days): Battery under load, pads,
internal electronics, full-energy charge cycle, Shock
button, and software.
Open Lid (when lid is opened): Battery, pads, internal
electronics, Shock button, and software.
Close Lid (when lid is closed): Battery, pads, internal
electronics, Shock button, and software.
Technical Data