
CAPINTEC, INC.
CAPTUS
®
600
1. WELL
2. PROBE
0. RETURN TO PREVIOUS SCREEN
14:32
CALIBRATION
Figure 6-9
Press
1
to select
WELL
or
2
to select
PROBE
. After making this selection, the appropriate
Calibration Screen will appear. Figure 6-10 is an example of the Probe Calibration Screen. If
data for a cesium standard was entered in the Cs137 Standard Setup, the source name will
appear on the display. Otherwise this field will be blank.
Note: Even if the CAPTUS
®
600 is equipped with only one detector used for both Probe and
Well measurements, separate Auto Calibration must be done for the Well and Probe
in this section.
1 4 :3 2
P R O B E
C A L IB R A T IO N
M E A S U R E C s 13 7 :
S T A R T
C S 1
Figure 6-10
Place the Cs137 standard source in the CAPTUS
®
600 detector. Press the
[F4]
START
key.
The counting display (Figure 6-11) will appear. As the counts increase, the vertical scale also
increases, changing from 100, to 200, 500, etc. to accommodate for the number of counts
measured by the CAPTUS
®
600. The CAPTUS
®
600 will count until sufficient counts are
accumulated for accurate calibration. After the AutoCal program sets the High Voltage, it
counts until the 662 keV gamma peak reaches 1000 counts. It then calculates the full width-
half maximum (FWHM) of the peak. This FWHM is used in calculating peak widths based on
the energy resolution of the detector. The system uses the FWHM calculated in AutoCal or
Test
,
whichever was done last.
November 07
ACCEPTANCE & QUALITY ASSURANCE TESTS
6-7
Содержание CAPTUS 600
Страница 2: ......
Страница 40: ......
Страница 62: ......
Страница 77: ...CAPINTEC INC CAPTUS 600 Page 1 November 07 ACCEPTANCE QUALITY ASSURANCE TESTS 6 15...
Страница 78: ...CAPINTEC INC CAPTUS 600 Page 2 Page 3 6 16 ACCEPTANCE QUALITY ASSURANCE TESTS November 07...
Страница 90: ......
Страница 102: ......
Страница 112: ...CAPINTEC INC CAPTUS 600 Figure 9 15 9 10 THYROID UPTAKE ONE CAPSULE METHOD November 07...
Страница 120: ...CAPINTEC INC CAPTUS 600 Figure 10 11 10 8 THYROID UPTAKE TWO CAPSULES METHOD November 07...
Страница 131: ...CAPINTEC INC CAPTUS 600 Figure 11 16 November 07 THYROID UPTAKE STANDARD CAPSULE METHOD 11 11...
Страница 132: ......
Страница 150: ...CAPINTEC INC CAPTUS 600 Figure 13 11 13 8 WIPE TEST November 07...
Страница 152: ......
Страница 162: ...CAPINTEC INC CAPTUS 600 Figure 15 5 15 4 LAB TESTS URINE TEST November 07...
Страница 167: ...CAPINTEC INC CAPTUS 600 Figure 16 5 November 07 LAB TESTS BLOOD TESTS 16 5...
Страница 171: ...CAPINTEC INC CAPTUS 600 Figure 16 9 November 07 LAB TESTS BLOOD TESTS 16 9...
Страница 175: ...CAPINTEC INC CAPTUS 600 Figure 16 14 November 07 LAB TESTS BLOOD TESTS 16 13...
Страница 176: ......
Страница 186: ......
Страница 208: ......
Страница 214: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......