Input the desired Counting Period; the maximum accepted value is 999 seconds, and the
minimum accepted value is 1 second.
X-RAY/GAMMA RATIO
These factors are necessary because the I125 X-Ray peak overlaps with the I131 X-Ray
peak and the I123 X-Ray peak. The CAPTUS
®
600 calculates I125 activity in Bioassay by
first searching for the I131 peak at 364 keV. If the I131 peak is detected, the contribution
from the I131 X-ray (counts in the I131 gamma peak times the I131 X-Ray gamma ratio) is
subtracted from the I125 X-Ray region of interest. It then searches for the I123 peak at
159keV. If the I123 peak is detected, the contribution from the I123 X-Ray (counts in the I123
gamma peak times the I123 X-Ray gamma ratio) is subtracted from the I125 X-Ray region of
interest. The default values are 8.5% for I131 and 92% for I123 and have been derived from
actual test data.
The last field is the DISTANCE, in centimeters, between the Probe and the neck. This must
be set to the same distance used in calculating Conversion Factors for the Probe for I125,
I123 and I131.
When all information is correct, press the
F4 [OK]
key.
Note: In order to report the Bioassay in units of activity rather than in counts, the CAPTUS
®
600 must have Conversion Factors for I125, I123 and I131 in the Probe in memory.
Reference CHAPTER5: SYSTEM INITIALIZATION, SECTION CONVERSION
FACTORS.
The
Bioassay Setup
menu (Figure 17-3) will re-appear.
Note: It is possible to perform a Bioassay without associated staff data. The results of such
a test will not be stored.
MAKING BIOASSAY MEASUREMENTS
In the Main Probe Screen, press the
F4 [BIO]
key. The Bioassay Initials Screen (Figure 17-
8) will appear:
November 07
STAFF BIOASSAY
17-5
Содержание CAPTUS 600
Страница 2: ......
Страница 40: ......
Страница 62: ......
Страница 77: ...CAPINTEC INC CAPTUS 600 Page 1 November 07 ACCEPTANCE QUALITY ASSURANCE TESTS 6 15...
Страница 78: ...CAPINTEC INC CAPTUS 600 Page 2 Page 3 6 16 ACCEPTANCE QUALITY ASSURANCE TESTS November 07...
Страница 90: ......
Страница 102: ......
Страница 112: ...CAPINTEC INC CAPTUS 600 Figure 9 15 9 10 THYROID UPTAKE ONE CAPSULE METHOD November 07...
Страница 120: ...CAPINTEC INC CAPTUS 600 Figure 10 11 10 8 THYROID UPTAKE TWO CAPSULES METHOD November 07...
Страница 131: ...CAPINTEC INC CAPTUS 600 Figure 11 16 November 07 THYROID UPTAKE STANDARD CAPSULE METHOD 11 11...
Страница 132: ......
Страница 150: ...CAPINTEC INC CAPTUS 600 Figure 13 11 13 8 WIPE TEST November 07...
Страница 152: ......
Страница 162: ...CAPINTEC INC CAPTUS 600 Figure 15 5 15 4 LAB TESTS URINE TEST November 07...
Страница 167: ...CAPINTEC INC CAPTUS 600 Figure 16 5 November 07 LAB TESTS BLOOD TESTS 16 5...
Страница 171: ...CAPINTEC INC CAPTUS 600 Figure 16 9 November 07 LAB TESTS BLOOD TESTS 16 9...
Страница 175: ...CAPINTEC INC CAPTUS 600 Figure 16 14 November 07 LAB TESTS BLOOD TESTS 16 13...
Страница 176: ......
Страница 186: ......
Страница 208: ......
Страница 214: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......