CAPINTEC, INC.
CAPTUS
®
600
SAVE
COUNT
SC/CR
5
400 keV
100 Cts FS 256 keV cc: 3 TOTAL
P
lt: 10.0 sec
RANGE
Figure 19-6
The function keys on this screen operate as follows:
•
F1 [COUNT]
Re-counts the background without saving the current
measurement.
•
F2 [RANGE]
Switches between the selected ranges. This key will be blank and
inactive if only one energy range was chosen.
•
F3 [SAVE]
Saves the displayed spectrum to memory. The word “SAVED” is
displayed.
•
F4 [SC/CR]
Used to either change the vertical scale of the display or the
number of channels the cursor will move for each press of an
ARROW
section on page 19-7 for
usage instructions.)
•
PRINT [P]
Prints the Background Measurement.
To exit Background
and measure Sources, press the
BSP
key. The Measurement Selection
Screen (Figure 19-3) will re-appear.
MEASURING A SOURCE
When
MEASURE SOURCE
is chosen, screens similar to MCA Setup will appear (except for
the name screen). If a Setup has been chosen, the values stored in the Setup will be
displayed. If no Setup has been chosen, the default values will be displayed. Even if a Setup
has been chosen, the values can be changed at this point, but they will only be valid for this
measurement and will not be stored.
The ROI Entry Screens will appear.
19-4
MULTICHANNEL ANALYZER OPERATION
November 07
Содержание CAPTUS 600
Страница 2: ......
Страница 40: ......
Страница 62: ......
Страница 77: ...CAPINTEC INC CAPTUS 600 Page 1 November 07 ACCEPTANCE QUALITY ASSURANCE TESTS 6 15...
Страница 78: ...CAPINTEC INC CAPTUS 600 Page 2 Page 3 6 16 ACCEPTANCE QUALITY ASSURANCE TESTS November 07...
Страница 90: ......
Страница 102: ......
Страница 112: ...CAPINTEC INC CAPTUS 600 Figure 9 15 9 10 THYROID UPTAKE ONE CAPSULE METHOD November 07...
Страница 120: ...CAPINTEC INC CAPTUS 600 Figure 10 11 10 8 THYROID UPTAKE TWO CAPSULES METHOD November 07...
Страница 131: ...CAPINTEC INC CAPTUS 600 Figure 11 16 November 07 THYROID UPTAKE STANDARD CAPSULE METHOD 11 11...
Страница 132: ......
Страница 150: ...CAPINTEC INC CAPTUS 600 Figure 13 11 13 8 WIPE TEST November 07...
Страница 152: ......
Страница 162: ...CAPINTEC INC CAPTUS 600 Figure 15 5 15 4 LAB TESTS URINE TEST November 07...
Страница 167: ...CAPINTEC INC CAPTUS 600 Figure 16 5 November 07 LAB TESTS BLOOD TESTS 16 5...
Страница 171: ...CAPINTEC INC CAPTUS 600 Figure 16 9 November 07 LAB TESTS BLOOD TESTS 16 9...
Страница 175: ...CAPINTEC INC CAPTUS 600 Figure 16 14 November 07 LAB TESTS BLOOD TESTS 16 13...
Страница 176: ......
Страница 186: ......
Страница 208: ......
Страница 214: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......