CAPINTEC, INC.
CAPTUS
®
600
13:06
OK
ENRGY
MULTICHANNEL ANALYZER
SETUP
ENERGY SCALE : 200 keV
COUNTING TIME :> 60 SEC X
COUNTING TIME BASE: REAL
BASE
CUT OFF CHANNEL : 5
PEAK COUNTS : 100
P/T
Figure 18-6
The MCA counting can stop when the counting time is reached or when the Peak Counts
(i.e. Full Scale Counts) reach a given value. Only one of these options is active. The “X”
indicates that Counting Time is active.
Input the Counting Time in seconds and Low-level Cutoff Channel. Any input pulses that are
measured below this cutoff channel will be ignored by the MCA. To calculate the cutoff in
keV, divide the desired cutoff in keV by the gain of the energy range selected. The gains for
the energy ranges are as follows:
200 keV scale .......... 1 keV/Channel
400 keV scale .......... 2 keV/Channel
800 keV scale .......... 4 keV/Channel
1600 keV scale .......... 8 keV/Channel
EXAMPLE:
For a 40 keV cutoff on the 400 keV full scale range, the desired cutoff channel
is calculated as:
20
=
Channel
/
keV
2
keV
40
=
channel
cutoff
Press
F1 [ENRGY]
to change the Energy Scale from 200 keV to 400 keV to 800 keV to 1600
keV.
Press
F2 [BASE]
to switch the Counting Time Base between REAL and LIVE (Reference
CHAPTER 3: GENERAL OPERATING OVERVIEW for an explanation of Real and Live
Time).
Press
F3 [P/T]
to switch between Peak Counts and Counting Time.
When Peak Counts is selected, the “X” will move from the Counting Time to the Peak Counts
as shown in Figure 18-7.
18-4
MULTICHANNEL ANALYZER SETUP
November 07
Содержание CAPTUS 600
Страница 2: ......
Страница 40: ......
Страница 62: ......
Страница 77: ...CAPINTEC INC CAPTUS 600 Page 1 November 07 ACCEPTANCE QUALITY ASSURANCE TESTS 6 15...
Страница 78: ...CAPINTEC INC CAPTUS 600 Page 2 Page 3 6 16 ACCEPTANCE QUALITY ASSURANCE TESTS November 07...
Страница 90: ......
Страница 102: ......
Страница 112: ...CAPINTEC INC CAPTUS 600 Figure 9 15 9 10 THYROID UPTAKE ONE CAPSULE METHOD November 07...
Страница 120: ...CAPINTEC INC CAPTUS 600 Figure 10 11 10 8 THYROID UPTAKE TWO CAPSULES METHOD November 07...
Страница 131: ...CAPINTEC INC CAPTUS 600 Figure 11 16 November 07 THYROID UPTAKE STANDARD CAPSULE METHOD 11 11...
Страница 132: ......
Страница 150: ...CAPINTEC INC CAPTUS 600 Figure 13 11 13 8 WIPE TEST November 07...
Страница 152: ......
Страница 162: ...CAPINTEC INC CAPTUS 600 Figure 15 5 15 4 LAB TESTS URINE TEST November 07...
Страница 167: ...CAPINTEC INC CAPTUS 600 Figure 16 5 November 07 LAB TESTS BLOOD TESTS 16 5...
Страница 171: ...CAPINTEC INC CAPTUS 600 Figure 16 9 November 07 LAB TESTS BLOOD TESTS 16 9...
Страница 175: ...CAPINTEC INC CAPTUS 600 Figure 16 14 November 07 LAB TESTS BLOOD TESTS 16 13...
Страница 176: ......
Страница 186: ......
Страница 208: ......
Страница 214: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......